Product Specifications (including optional items)
[Optional Accessory] How should I consider the selection between capacitor coupling and arrester coupling for LSS-INJ6401SIG?
LSS-INJ6401SIG can switch between capacitor coupling and arrester coupling, but with capacitor coupling, communication may not be possible because the impedance changes. Arrester coupling can be selected between "Capacitor/Arrester" and "Arrester ...
[Optional Accessory] Regarding LSS-INJ6401SIG, there are three types: capacitor coupling, gas arrester coupling, and gas arrester + capacitor coupling. Please explain the criteria for selecting each.
・Capacitor coupling Compliant with IEC 61000-4-5 Ed.3. Since it is not a switching element, the waveform rise is clean. However, because of the large capacitance C, impedance drops significantly, making AE-EUT communication unlikely to succeed, and ...
[Test Equipment] What are the dimensions and weight of the LSS-F03 series lightning surge tester?
The dimensions and weight are as follows: Dimensions (excluding protrusions): LSS-F03A1A/A3A: W555×H1450×D790 (mm), LSS-F03C1A/C3A: W555×H1800×D790 (mm) Weight: A1A: Approximately 290kg, A3A: Approximately 300kg, C1A: Approximately 325kg, C3A: ...
[Test Equipment] What are the dimensions and weight of the LSS-6330 series lightning surge tester?
The dimensions and weight are as follows: Dimensions (excluding protrusions) LSS-6330-A20A: W430×H349×D540 (mm) LSS-6330A-B63A: Surge generator section: W430×H371×D540 (mm), CDN section W430×H695×D686 (mm) Weight: LSS-6330-A20A Approximately 50kg, ...
[Test Equipment] What happens when EUT power capacity exceeding the rated value is input?
If voltage and current exceeding the rated values are input, it may cause damage to the tester or fire. It is conceivable that CDN components or wiring inside the tester may disconnect due to heat, so please use within the rated specifications.
[Test Equipment] Is there a specified warm-up time for lightning surge testers?
There is no specified warm-up time.
[Test Equipment] What is the power consumption of lightning surge testers?
The power consumption of each tester is as follows: LSS-F03 series: Approximately 400VA LSS-6330-A20: Approximately 230VA ? LSS-6330A-B63A Surge generator section: Approximately 230VA LSS-6330A-B63A CDN section: Approximately 85VA
[Test Equipment] Does the LSS-F03 series comply with IEC 60664? (Dielectric withstand test)
It does not comply because the waveform interpretation and peak voltage tolerance are different. IEC 60664: Peak voltage tolerance is ±3% IEC 61000-4-5: LSS-F03 has a specification of ±10% relative to the set voltage IEC 60664: Uses IEC61000-4-5 ...
[Test Equipment] The specification section in the instruction manual for the LSS-F03 series lightning surge tester states "Residual voltage: 15% or less of the applied surge voltage, or twice or less the peak value of the rated voltage of the power line."
After testing, no voltage remains because the discharge circuit inside the tester operates. This specification item in the instruction manual measures the voltage at the input section while applying, with the input and output sections open. This is a ...
[Test Equipment] Regarding the discharge of lightning surge testers LSS-6330 series and LSS-F03 series, the instruction manual states <When the test ends, a discharge time is provided for the internal high-voltage circuit. The discharge time is approximat
The following situations require approximately 4 seconds for discharge to complete: "When test ends" "When test stops (STOP or ERROR)" During "emergency stop," forced discharge is performed, so there is no time constraint. Also, during "test pause," ...
[Optional Accessory] Please explain the difference between unshielded unbalanced lines and unshielded balanced lines when selecting between LSS-INJ6401SIG and F-130814-1004.
[What is an unbalanced line] Communication where signals are transmitted through voltage fluctuation based on a single 0V line as reference Example: RS-232C (communication between devices), etc. [What is a balanced line] Communication where signals ...
[Optional Accessory] Is there a method to conduct testing while communicating via RS-485 (balanced line) using LSS-INJ6401 or Telecom CDN?
Testing RS-485 while communicating via LSS-INJ6401 or Telecom CDN is difficult. LSS-INJ6401 is designed for unbalanced line testing, and since the decoupling L is not wound in common mode, differential signals are unlikely to pass through. The ...
[Optional Accessory] Can testing be performed for each of the following LAN cable specifications/specs ①-④ with the high-speed communication dedicated superposition unit MODEL: F-130814-1004?
①CAT5E UTP (unshielded) ②CAT5E STP (shielded) ③CAT6 UTP (unshielded) ④CAT6 STP (shielded) ? ? Since CAT5E and CAT6 have a maximum communication speed of 1000Mbps, testing with F-130814-1004 is possible. ①CAT5E UTP (unshielded) and ③CAT6 UTP ...
[Optional Accessory] Please tell me the capacity of the TF-6503P isolation transformer.
It is 52 kVA. For a three-phase isolation transformer, calculate using the following formula: [Maximum input voltage 600V] × [Maximum output current 50A] × √3 = 52kVA
[Test Equipment] I am using a custom-made lightning surge tester that can turn OFF AC LINE detection. I turned OFF AC LINE detection to check the waveform at AC LINE OUT, but a line detection error occurred and application is not possible. Are there any o
In addition to AC LINE detection, please set the phase angle to asynchronous in MANUAL mode and set EUT LINE SW to ON.
[Test Equipment] Why is the waveform during power superposition not specified in the instruction manual for the JEC standard-compliant LSS-700 series?
This is because the JEC standard only specifies waveforms for surge output and does not include descriptions for power superposition. Therefore, even if you observe the waveform at CDN OUT of the LSS-700 series, the voltage waveform will not be ...
[Test Equipment] In the LSS-700 series, a protective transformer is installed on the secondary side of the EUT line, but back surge is observed. Is it possible to further attenuate this back surge voltage?
We believe it is difficult to further suppress the back surge. ・When adding an isolation transformer: Since it is an isolation transformer, normal mode surge will pass through. ・When inserting an LC filter: It is possible to further reduce the back ...
[Optional Accessory] 18-00048B has a label indicating 4.5kV compatibility on the outlet section. Is it not possible to conduct tests at voltages higher than this?
The dielectric withstand voltage of the outlet section is 4.5kV, but the dielectric withstand voltage of the terminal section is 15kV. Therefore, for tests above 4.5kV, please wire to the terminal and conduct the test.
[Optional Accessory] Please tell me the manufacturer and model number of the outlet section for 18-00048B.
It is manufactured by Panasonic Corporation, model WCF5901. https://www2.panasonic.biz/ideacontout/2012/12/08/2012120800030113.PDF
[Optional Product] Can a step-up transformer be used as a substitute for the isolation transformer used in lightning surge testers?
There is no problem as long as it is an isolation transformer. However, general step-up transformers and variable autotransformers (slidacs) have PE connected between the primary and secondary sides, and are not completely isolated, so they cannot be ...
[Optional Product] Please tell me the capacity of the TF-2302P isolation transformer.
7.2 kVA. Calculated using the following formula: Drive power supply [Maximum input voltage 240V] × [Maximum output current 30A] = 7.2 kVA
[Main Unit] The guaranteed current surge output range of the LSS-720B2 is 800 A to 4000 A. What kind of waveform is produced when output is below 800 A?
There is a possibility that a waveform with the undershoot shifted later may be observed.
[Test Equipment] Does a lightning surge test need to be conducted in a shielded room?
A shielded room is not required to conduct a lightning surge test. This does not mean that a shielded room must not be used; many facilities conduct the test in a shielded room. When conducting the test, be sure to insulate the EUT and cables from ...
[Test Equipment] Does LSS-720B2 comply with JEC-0203 and JEC-0204?
LSS-720B2 does not comply with JEC-0203 and JEC-0204. Custom modifications are also not available. The following are the points where compliance cannot be achieved according to each standard. ●JEC-0203 "○7.4.4 Standard lightning impulse voltage ...
[Test Equipment] Can lightning surge tests be performed on an EUT that has voltage at the application point (e.g., output port of power supply equipment, power conditioner, etc.)?
・LSS-6330: Testing is possible if it does not exceed the CDN voltage. Phase angle application is possible by turning LINE SW ON. Refer to the manual "EUT LINE Voltage Detection" (Application from pulse out is not possible) ・LSS-F03: Software ...
[Test Equipment] I am using the LSS-15AX-A1 and LSS-F03-A1 lightning surge testers. When applying to the phase angle of the AC line, the 15AX manual describes phase angle correction, but there is no description for the F03. Is correction not necessary for
The F03 automatically corrects based on the application phase, so there is no need to correct and input the phase angle. Supplement LSS-15AX: Manual correction LSS-F02/F03: Automatic correction LSS-F03A: Since around October 2019, automatic ...
[Test Equipment] Are there any consumable parts?
For LSS-F03series, the changeover switch, charging capacitor, discharge relay, discharge switch, contactor, breaker, tube fuse, fan, varistor, and arrester used inside the main unit are consumable parts. The timing of parts replacement varies ...
[Optional Accessory] Please tell me the lifespan of the gas arrester used in the communication line coupling CDN for lightning surge testers.
The lifespan of gas arresters varies greatly depending on the set voltage, so we cannot provide a uniform lifespan. Regardless of the number of applications, the arrester needs to be replaced when it no longer emits light when a surge is applied. The ...
[Optional Accessory] Please tell me about the compatibility of connection cables between the LSS-F03 series and external signal line CDN. High-speed communication line coupling unit (F-130814-1004) - For F-130814-1004: 05-00147A, Interconnection dedicated
The 05-00147A used with F-130814-1004 has a cable length of 1.5m to meet the standard values when calibrated in combination with the LSS-F03. The 05-T1578 used with LSS-INJ6401SIG has a cable length of 2m, assuming the unit will be placed on the ...
【Tester】Can the LSS-F03C type perform tests to the IEC 62368-1 standard?
IEC 62368-1 specifies the circuit for the impulse generator, so the 1.2/50 generator differs from the specified circuit. (The 10/700 circuit is the same.) Therefore, the standard LSS-F03C type cannot perform tests to the IEC 62368-1 standard.
[Test Equipment] What voltage is the standard power cable rated for?
The standard power cable is for AC100-120V. When using AC220-240V, please prepare a 3-conductor power cable with a protective ground terminal that complies with the safety standards of the country of use.
【Option】What is the operating voltage of the 11-00017A/11-00008B warning light?
The warning light is rated at DC 24 V. It is operated through its connection to the main unit.
[Test Equipment] How should I replace the fuse on the lightning surge tester?
The LSS-F03series has a fuse holder in the AC inlet on the front panel, and the LSS-6330series has one on the rear panel, which can be replaced by the customer. If the recommended fuse is difficult to obtain, please contact our Customer Service ...
[Test Equipment] Why did the saved test conditions disappear after not using the lightning surge tester for a while?
Some testers have an internal battery that charges when connected to the drive power supply. If the tester is not connected to power for a while, the internal battery discharge progresses, and the test conditions saved in the tester may be deleted. ...
[Test Equipment] Is it acceptable to conduct testing (line surge or CDN test) while the L1, L2, L3, N, PE output line cables (red) and the HOT, COM output line cables (red) to CDN are connected to the LSS-F03 series lightning surge tester and connected to
In that case, an ERROR will be displayed and testing cannot be performed. When attempting to conduct a test using the surge output terminals [HOT] and [COM] of the tester... ERROR4 When attempting to conduct a test using the AC/DC line coupling ...
【Tester Main Unit】Can the LSS-F03 series perform normal-mode testing on telecom lines?
The LSS-F03 series cannot perform normal-mode testing on telecom lines. This is because IEC61000-4-5.Ed3 clarified the test requirements for telecom lines and specified common mode only. ? During the IEC61000-4-5.Ed1 era, when the LSS-15AX series was ...
[Test Equipment] What is the waveform guarantee when using the LSS-F series lightning surge tester (variable choke coil) with each choke coil (0.8/1.0/1.3/1.5mH)?
The waveform guarantee applies only to 1.5mH. The waveform will change when other choke coils are selected. The purpose of using choke coils other than 1.5mH is to shift the resonance point for products such as high-current products where evaluation ...
[Optional Accessory] What are the system requirements for the LSS-F03 control software? (OS, CPU, Memory)
There is no problem as long as the system meets the following requirements: OS: Win10/11, CPU: 1GHz or faster processor or SoC, Memory: 1GB (32-bit) or 2GB (64-bit).
[Optional Accessory] When inputting AC100V to the isolation transformer TF-2302P and checking the output side L1-PE and L2-PE with a tester, the readings vary between 70V and 30V each time. Is this normal?
This is normal. When measuring the secondary side of the isolation transformer without connecting anything, there is no reference because it is floating from ground. Please verify that 100V is measured between L1-L2.
[Optional Accessory] The isolation transformer TF-6503P for lightning surge testing has separate taps for 240V, 480V, and 600V for each input/output voltage. Is it acceptable to connect and use low voltages such as 100V to the 600V terminal?
When connecting a low voltage to a high voltage tap, the voltage fluctuation from no-load to maximum load may exceed 5%. If voltage fluctuation cannot be tolerated, please switch to a tap with a closer value.