[Test Equipment] Regarding the discharge of lightning surge testers LSS-6330 series and LSS-F03 series, the instruction manual states <When the test ends, a discharge time is provided for the internal high-voltage circuit. The discharge time is approximat

[Test Equipment] Regarding the discharge of lightning surge testers LSS-6330 series and LSS-F03 series, the instruction manual states <When the test ends, a discharge time is provided for the internal high-voltage circuit. The discharge time is approximat

The following situations require approximately 4 seconds for discharge to complete:

  • "When test ends"

  • "When test stops (STOP or ERROR)"

During "emergency stop," forced discharge is performed, so there is no time constraint.
Also, during "test pause," only the count stops and discharge is not performed.