We believe it is difficult to further suppress the back surge.
・When adding an isolation transformer: Since it is an isolation transformer, normal mode surge will pass through.
・When inserting an LC filter: It is possible to further reduce the back surge, but this will result in increased leakage current.
When using CVCF as the EUT power supply, please check the dielectric withstand voltage. Also, if the dielectric withstand voltage is insufficient, please install surge protection elements externally.
(Back surge of up to approximately 2kV may occur with 20kV model testers, and approximately 1.2kV with 12kV models)