Product Specifications (including optional items)
[Optional Product] I believe that the isolation transformer TF-6503P can be used as a low-voltage transformer with 480V input → 240V output. May I also use the same terminal as 240V → 100V? (I anticipate high losses)
As for the structure of the isolation transformer, 480V terminal: 200V input / 240V terminal: 100V output is possible. However, since the specification of TF-6503P assumes that input and output are taken from the same terminals, there is no guarantee ...
[Option] Can the TF-6503P/TF-6633P be connected to single-phase three-wire wiring?
Yes. Please refer to the attached document.
[Test Equipment] Is it possible to set L+N -PE simultaneous application with the lightning surge tester LSS-F03 series?
It is possible with custom modifications to the lightning surge tester (LSS-F03 series) main unit and custom external CDN. We have experience with modifications for ITU-T K.21 K.44 standards compliance.
[Test Equipment] The lightning surge tester LSS-6230A complies with IEEE C62.45-2002, and the lightning surge tester LSS-6330 complies with IEC61000-4-12. What is the difference between these two?
IEEE C62.45-2002 does not have current waveform rise time specifications, while IEC61000-4-12 does have such specifications. Therefore, IEC61000-4-12 covers IEEE C62.45-2002.
[Test Equipment] Please tell me the capacitor capacity when outputting each waveform with the lightning surge tester LSS-15AX/F02/F03 series.
・When outputting 1.2/50μF combination waveform from the surge out section: 10μF ・When outputting 1.2/50μF combination waveform from the power superimposition section: 10μF+2μF=12μF ・When outputting 10/700μF combination waveform: 10μF+2μF+8μF=20μF
Confidential: [Test Equipment] Does the lightning surge tester LSS-720B2 support JEC0202?
It is supported. JEC0202 (1994) is the successor standard to JEC212 (1981). ⇒Changed to confidential as JEC0202 is not compliant. (2025.04.16)
[Optional Product] Which LAN cable categories does the high-speed communication line unit F-130814-1004 support?
It supports CAT5, CAT5e, CAT6, and CAT6A.
[Optional Product] Is the internal coupling method of the high-speed communication line unit F-130814 series capacitor or gas arrester coupling?
It is gas arrester coupling.
[Optional Product] Please tell me how to use the POE switch of the high-speed communication line unit F-130814 series.
The POE switch is used to avoid communication errors caused by resonance with the CDN during POE communication between the EUT and AE. When the switch is turned ON (green lamp lights up), the decoupling circuit is disconnected. Therefore, during ...
[Optional Product] I would like to conduct LAN evaluation with the mutual communication line unit. Is this possible?
Because the inductance is as high as 20mH and communication is not possible, it cannot be supported. Evaluation with the high-speed communication line unit F-130814-1004 is required.