Test method
Q. With the LSS-F03, can voltage be applied from CDN OUT to the EUT live line without supplying power to the EUT through the CDN?
Voltage cannot be applied to the EUT live line without supplying power through the CDN. If the test is started without supplying power, current will flow from the live line to the tester when the switching unit operates, potentially damaging the ...
We use the common ground in our factory. Is it necessary to install a separate ground (isolated ground) for lightning surge testing?
There are no specific grounding requirements. A common ground is acceptable.
Which is better suited for withstand voltage (insulation) testing, the lightning surge tester LSS-720 series or the F03 series (IEC standard)?
Follow the test requirement standards, but for withstand voltage (insulation) testing, LSS-720 is optimal.
I am applying surge to the EUT signal line through a varistor from SURGE OUT in lightning surge testing. What kind of varistor should be used?
If there are no specifications in the reference standard, selection is made based on the test level and impact on signal lines. Please confirm the following with multiple varistors and make your selection. (1) Confirm signal connectivity and waveform ...
If the EUT power supply does not have PE but has an FG terminal, where should the FG terminal be wired?
Follow the reference standard, but if there are no specifications, conduct normal mode testing without connection. However, considering the actual installation situation, it is preferable to also conduct common mode testing by connecting to the PE of ...
Is it necessary to wear insulating gloves when using a lightning surge tester? Does this equipment fall under regulations such as Article 341 of the Industrial Safety and Health Regulations regarding insulating protective equipment?
Insulating gloves are not mandatory. The instruction manual states "Turn off the power of this tester when making connection changes to cables and connection bars," so this is equipment that should "not be operated while live." There is no problem ...
How should the selection between capacitor coupling and arrester coupling be considered for LSS-INJ6401SIG?
In IEC61000-4-5 Ed.2, the coupling method was to be determined by higher-level standards or product specifications. However, as an announcement, there is a statement that capacitor coupling is the preferred method, and arrester coupling should be ...
Are qualifications required for using a lightning surge tester?
There is no need to obtain national qualifications or certified qualifications. However, the following important safety matters are written in most tester instruction manuals.
Is there a method to conduct RS-485 (balanced line) testing while communicating using LSS-INJ6401 or Telecom CDN?
It is difficult to conduct RS-485 testing while communicating using LSS-INJ6401 or Telecom CDN. LSS-INJ6401 is for asymmetrical line testing, and since the decoupling L is not wound in common mode, differential signals are not expected to pass ...
In lightning surge testing, is it possible to test an EUT that has voltage at the application point (e.g., output port of power supply equipment, power conditioner, battery, etc.)?
・LSS-6330 series: Testing is possible as standard if CDN voltage is not exceeded. Phase angle application is possible by turning LINE SW ON. ・LSS-F03 series: Firmware modification is required. After modification, the CDN voltage must not be exceeded. ...
Please tell me about the equipment arrangement (test environment).
The test environment for power lines involves placing an insulation support stand with a height of 0.1m on the ground plane, and placing the Equipment Under Test (EUT) on top of it. The length of the power line is 0.5m, or 1m for floor-standing ...
Is an isolation transformer required for lightning surge testing?
It is required when connecting AC power for EUT to the tester. If AC power is supplied directly to the tester, the leakage current breaker installed on the power supply may activate due to the influence of leakage current inside the tester, so please ...
Can a regulated power supply be used for the EUT AC line power supply?
Basically, there should be no problem. However, if the output of the regulated power supply is connected to the isolation transformer via a line filter, resonance may occur between the line filter and isolation transformer, causing waveform ...
(Confidential) I want to apply lightning surge to the DC line, but is an isolation transformer required?
Basically, it is not required. The DC superimposition section also has a decoupling circuit, so the primary side is protected, but even at maximum test voltage of 15kV normal mode application, approximately 200V line-to-ground voltage is applied to ...
In IEC 61000-4-5 ed.3, why is 10/700μs surge used for outdoor communication lines over 300m?
When there is a long transmission line, the surge coming through it is expected to have a slow pulse and small energy waveform due to the L component of the wiring. The 10/700?s simulates this and serves to complement the 1.2/50μs test. Therefore, ...
Please explain how to protect auxiliary equipment from back surge when superimposing surge onto input from AE while the EUT is powered by DC power supply.
The DC superimposition section also has a decoupling circuit, so the primary side is protected, but even at maximum test voltage of 15kV normal mode application, approximately 200V line-to-ground voltage is applied to the primary side. For 24V DC ...
When conducting a test on the shield connection line for lightning surge testing, is there any risk of electric shock if the tester stands on the ground plane?
There is no problem even if the operator is on the ground plane. This is because no circuit is formed for current to flow through the human body.
(Confidential - Duplicate) Is grounding work required when conducting tests?
Grounding work based on the Electrical Facilities Technical Standards is broadly classified into 4 types from A to D, but specifically, no grounding work is required for testing. However, please connect the power cord to a power outlet equipped with ...
When conducting a power line superimposition test with a lightning surge tester, the phase voltage output from the CDN becomes unstable when pressing the start button.
In the LSS series, the surge generation section also acts as a load. When the surge generation section (load) is connected to the application phase, depending on the AC power supply used, voltage drop in the application phase or voltage rise in the ...
When switching the application polarity between +/- without changing the injection & return phases on a lightning surge tester, versus switching the injection & return phases without changing the application polarity, will the noise waveform applied to th
It is considered that the noise waveform applied to the EUT will be the same.
(Confidential-Duplicate) The EUT is a DC battery and has voltage on the application side. How should lightning surge testing be conducted?
For LSS-F03 series, if the hot switch circuit does not operate, the current from the battery will burn the switching unit, so a paid modification "to enable LINE_SW ON at any time" is required. LSS-6330 series (discontinued) uses general relays for ...
Is excessive voltage applied when switching between positive and negative polarity in surge tests? (EUT: Capacitor)
Excessive voltage is not applied during polarity switching. Even if the EUT is a capacitor, surge energy is discharged due to the impedance of the generator circuit. However, for extremely large capacitance, it is recommended to discharge before ...
A common ground is used in the factory. Is it necessary to install a separate ground (isolated ground) for lightning surge testing?
There are no specific grounding requirements. A common ground may be used.
Does lightning surge testing need to be performed in a shielded room?
A shielded room is not required to perform lightning surge testing. This does not mean a shielded room must not be used; many facilities perform the test in one. When performing the test, ensure that the EUT and cables are insulated from the RGP.