Product Specifications (including optional items)
[Main Unit] ISS-7820 and ISS-7821 require a 200V power supply for pulses 5a and 5b. When conducting tests other than pulses 5a and 5b, can testing be performed without supplying 200V power?
When conducting tests other than 5a and 5b, testing can be performed using only a 100V power supply. Because the control circuitry operates on 100V, tests for 5a and 5b require a 100V supply in addition to 200V.
When I checked the attenuation of the No Load waveform verification ATT MODEL: 00-00007A with a network analyzer, it was 34?35dB rather than 40dB. Is it defective?
Because the input Z of the No Load waveform verification ATT (MODEL:00-00007A) is as high as 2.475kΩ, connecting it directly to a 50Ω system changes its characteristics due to reflection. When we check transmission characteristics using a network ...
[Test Equipment] Regarding Pulse2a of ISS-7610, it states tr=1+0/-0.5μs, but ISO7637-2 (2004) states tr=10+0/-0.5μs. *ISO7637-2 (2011) is tr=1+0/-0.5μs. In this case, will tr=10+0/-0.5μs be output when selecting ISO7637-2 (2004) on ISS-7610?
Even when ISS-7610 (2004) is selected, tr=1+0/-0.5μs is output. Based on the following, we interpret tr=10μs in ISO7637-2 (2004) as a typographical error. (Standard publication history) ISO7637-2 (1990) tr=1μs or less ISO7637-2 (2004) states ...
[Test Equipment] When performing tests by outputting 3a, 3b from ISS-7690 via remote control from PC with the ISS-7600 series, the test stops due to a communication error. How can I avoid the communication error?
The cause is thought to be that noise gets on the cable because the frequency component contained in the 3a, 3b pulses is high. Please take measures such as attaching multiple ferrite cores to the USB cable connecting ISS-7690 and PC, shortening the ...
[Test Equipment] Please tell me the specifications of the relay used in ISS-T1551.
The relay model is MR82BL-081. The manufacturer is MITHUBA (JIDECO).
[Test Equipment] Is it possible to control ISO7630 with ISS7601 and perform ISO 7637-3 CCC method testing?
Yes, it is possible if you set UA (battery voltage) to 0V and input US (test voltage). However, as with standalone control, we recommend starting the test with the DC output line switch OFF for safety reasons. (Even with UA set to 0V, if the DUT ...
[Optional Accessory] I want to upgrade the OS to Windows11. Is ISS-7601 (remote control software) compatible with Windows11?
An upgrade to ISS-7601 version 14-00056A (Win10, 11 compatible) is possible. For BP4610, versions 2.0 and later of the main unit are also compatible with Win10 and 11.
[Test Equipment] Pulse4, which was specified in ISO7637-2 2004 edition, was transferred to ISO16750-2 2012 edition as "Startup Profile Test" from the 2011 edition. Why can't the ISS-7600 series and ISS-7800 series perform the "Startup Profile Test"?
When transitioning from Pulse4 of ISO7637-2 to the "Startup Profile Test" of ISO16750-2 2012 edition, a sine wave was added at t8. The BP4610 unit that generates Pulse4 has low waveform generation capability, so it cannot generate complex waveforms ...
[Main Unit] Can SG-7040 comply with ISO16750-2(2023)?
Except for the following three tests, testing is possible by having the customer create waveform files. 4.4 AC voltage superimposition test Because the SG-7040 frequency setting is limited to 150kHz, it cannot meet the required 200kHz. 4.6.4 Load ...