[Main Unit] ISS-7820 and ISS-7821 require a 200V power supply for pulses 5a and 5b. When conducting tests other than pulses 5a and 5b, can testing be performed without supplying 200V power?

[Main Unit] ISS-7820 and ISS-7821 require a 200V power supply for pulses 5a and 5b. When conducting tests other than pulses 5a and 5b, can testing be performed without supplying 200V power?

When conducting tests other than 5a and 5b, testing can be performed using only a 100V power supply.
Because the control circuitry operates on 100V, tests for 5a and 5b require a 100V supply in addition to 200V.