[Test Equipment] When performing tests by outputting 3a, 3b from ISS-7690 via remote control from PC with the ISS-7600 series, the test stops due to a communication error. How can I avoid the communication error?

[Test Equipment] When performing tests by outputting 3a, 3b from ISS-7690 via remote control from PC with the ISS-7600 series, the test stops due to a communication error. How can I avoid the communication error?

The cause is thought to be that noise gets on the cable because the frequency component contained in the 3a, 3b pulses is high.
Please take measures such as attaching multiple ferrite cores to the USB cable connecting ISS-7690 and PC, shortening the USB cable, or ensuring proper PC grounding.