[Test Equipment] Regarding Pulse2a of ISS-7610, it states tr=1+0/-0.5μs, but ISO7637-2 (2004) states tr=10+0/-0.5μs. *ISO7637-2 (2011) is tr=1+0/-0.5μs. In this case, will tr=10+0/-0.5μs be output when selecting ISO7637-2 (2004) on ISS-7610?

[Test Equipment] Regarding Pulse2a of ISS-7610, it states tr=1+0/-0.5μs, but ISO7637-2 (2004) states tr=10+0/-0.5μs. *ISO7637-2 (2011) is tr=1+0/-0.5μs. In this case, will tr=10+0/-0.5μs be output when selecting ISO7637-2 (2004) on ISS-7610?

Even when ISS-7610 (2004) is selected, tr=1+0/-0.5μs is output.

Based on the following, we interpret tr=10μs in ISO7637-2 (2004) as a typographical error.
(Standard publication history)
ISO7637-2 (1990) tr=1μs or less
ISO7637-2 (2004) states tr=10+0/-0.5μs.
It was later corrected to tr=1+0/-0.5μs in the amended version ISO7637-2 (2007/DIS).
Other test equipment manufacturers also output tr=1+0/-0.5μs for ISO7637-2 (2004).