[Test Equipment] Pulse4, which was specified in ISO7637-2 2004 edition, was transferred to ISO16750-2 2012 edition as "Startup Profile Test" from the 2011 edition. Why can't the ISS-7600 series and ISS-7800 series perform the "Startup Profile Test"?

[Test Equipment] Pulse4, which was specified in ISO7637-2 2004 edition, was transferred to ISO16750-2 2012 edition as "Startup Profile Test" from the 2011 edition. Why can't the ISS-7600 series and ISS-7800 series perform the "Startup Profile Test"?

When transitioning from Pulse4 of ISO7637-2 to the "Startup Profile Test" of ISO16750-2 2012 edition, a sine wave was added at t8.
The BP4610 unit that generates Pulse4 has low waveform generation capability, so it cannot generate complex waveforms like sine waves and requires an external signal generator.
Therefore, please use the DC fluctuation tester (SG-7040A) for the "Startup Profile Test".