Is aging of the test instrument required before conducting an ESD test?
The standard states that, unless otherwise specified, the climatic conditions for testing must be within the operating limits specified by the respective manufacturers for the EUT and test equipment.
Are there any precautions when using the discharge gun CR unit (500pF-0Ω)?
Because the energy is large and more current flows, the insulation performance inside the high-voltage switch deteriorates faster than during normal use, causing leakage. As a result, voltage is generated at the discharge electrode even in a charged ...
Will differences in CR unit values affect test results?
For capacitors, if the capacitance (unit: F) increases, the energy amount increases, making it more severe in terms of energy. For resistors, if the resistance value (unit: Ω) is small, current flows instantaneously, and if large, current flows ...
When should the EUT be discharged?
Discharge should be performed before each application. If the EUT is charged, the difference from the test voltage of the discharge gun becomes smaller, and the expected discharge current will not flow.
Please tell me about the test voltage when conducting air discharge testing under IEC 61000-4-2 Ed.2.
In the case of air discharge testing, the standard specifies that the designated test level and all lower levels should be performed. The reason is that at low levels (e.g., 2kV) and high levels (e.g., 8kV), the discharge destination of the air ...
Is it okay to use rubber gloves to prevent electric shock?
You may use rubber gloves. In IEC 61000-4-2, which is the standard for electrostatic testing, waveform verification is performed without human intervention, so during testing, either bare hands or gloves are acceptable. There is no specification for ...
Is grounding work required when conducting tests?
Grounding work based on the Technical Standards for Electrical Facilities is broadly classified into four types, A through D, but grounding work is not required when conducting tests. However, please connect the power cord to a power outlet equipped ...
Is there any problem with applying to terminals outputting DC24V or AC100V?
There is no problem. Do not connect the ground cable of the discharge gun to the EUT; connect it to the ground plane as usual. Connecting it to the EUT may cause short circuits or damage to the power supply of the electrostatic tester.
Since it is difficult to hold the discharge gun grip with one hand for a long time, is it okay to hold other parts of the discharge gun during testing?
Please conduct the test by holding the grip with one hand. If you support other parts of the discharge gun body or discharge cup with your other hand other than the grip, the output waveform may change (Ip increase) due to human influence ...
What are direct application (direct discharge) and indirect application (indirect discharge)?
Direct application (direct discharge) is a method of applying contact discharge or air discharge to the EUT, and indirect application (indirect discharge) is a method of applying by contacting a conical discharge tip to a horizontal coupling plate or ...
What are contact discharge and air discharge?
Contact discharge applies by contacting the metal parts of the EUT, and air discharge applies by approaching the non-metal parts of the EUT. These are differences in operation modes. Contact discharge uses a conical discharge tip, and air discharge ...