Inspection (Pre-shift inspection)
Q. I would like to check the waveform. How should this be done?
In addition to the impulse noise tester, prepare an oscilloscope (frequency bandwidth DC to 600 MHz or higher) and the waveform observation attenuator 00-00017A, and check the waveform output from PULSE OUT of the impulse noise tester by passing it ...
Q. Can the waveform observation attenuator 00-00017A be used to observe 4 kV or 2 kV waveforms?
The maximum voltage in the product specifications is the voltage when terminated with 53.5Ω. Therefore, when terminated with 50Ω, the maximum value is 3.96 kV for the 4 kV version (1.98 kV for the 2 kV version).
When the waveform of INS-4040 was measured, the rise time was 1.9 ns. Is there a reason why it became slower?
The cause is considered to be the low bandwidth of the ATT and oscilloscope. * 00-00017A: Since it is DC to 2 GHz, it can be checked at 1 ns or less.
The waveform observed at PULSE OUT is within specification, but when observed at EUT LINE OUTPUT it does not meet the specification. Is the tester defective?
The impulse noise tester has a 50Ω output impedance system. When measuring at the pulse output, measurement is performed using a 50Ω coaxial cable and a 50Ω-input attenuator (00-00017A), so measurement can be made under a fully matched condition, and ...
Q. When checking the applied waveform with a semiconductor impulse tester, an approximately 120 MHz (8.3 ns) damped oscillation waveform is observed after the rise. The waveform was measured at CDN OUT using the probe supplied with the oscilloscope. What
This is a characteristic of the semiconductor switch being used.
How should I perform waveform verification?
Please prepare an oscilloscope (frequency bandwidth DC to 600MHz or higher) and a waveform observation attenuator 00-00017A in addition to the impulse noise tester, and verify the waveform output from the PULSE OUT of the impulse noise tester through ...