Standard
Does the recommended test level vary depending on the DC supply voltage (12 V/3.3 V, etc.)?
In general, we recommend using a test level equivalent to that for control and signal lines (e.g., Level 1: 0.5 kV) as the basis for consideration. Most DC power inputs are on the secondary side of an AC/DC converter. Because external noise is ...
Is there a reason why the evaluation time for the EFT/B test is 1 minute?
Bursts are applied at a repetition frequency of 5kHz or 100kHz. Since burst pulses are not applied constantly, and depending on timing, they may not be applied at the point where a malfunction occurs, bursts are applied for 1 minute to apply them at ...
Please explain the climatic conditions in IEC 61000-4-4 Ed.3.
Unless specifically specified by the committee responsible for the generic standard or product standard, the climatic conditions are within the limits specified by each manufacturer for operation of the EUT and test equipment. * In IEC 61000-4-2, ...
What is the IEC standard number in which the fast transient/burst test is specified?
It is specified in IEC 61000-4-4.
Please explain the transition (changes) from Edition 1.0 to Edition 3.0 of the IEC 61000-4-4 standard.
Changes have been made to the application method for power lines and the CDN calibration method.
The CDN specifications for the FNS-AX3 simulator compliant with IEC 61000-4-4 Ed2 Amd1 are rise time 5ns ±30% and pulse width 50ns ±30%, while the CDN specifications for the FNS-AX4 simulator compliant with Ed3 are rise time 5.5 ±1.5ns and pulse width 45
Each simulator is described based on the standard document in effect at the time of manufacture, but the CDN design is almost unchanged. Therefore, although there are individual differences, there is no major difference in the waveform. ...
Please explain why the waveform specification at the CDN output section in IEC 61000-4-4 Ed3.0 became a rise time of 5.5ns and a half-value width of 45ns (whereas Ed2 had a rise time of 5ns and a half-value width of 50ns).
The waveform specification was changed in consideration of the wiring length (inductance) from the generator section to the CDN section.
Why are the specified distances between the EUT and the simulator different for tabletop equipment (0.5m) and floor-standing equipment (1.0m)?
We assume the difference in length is due to the size of the EUT equipment. Floor-standing equipment is likely larger than tabletop equipment, so the cable length is considered to be longer for floor-standing equipment.
Please tell me about the transition of the standards.
The IEC 61000-4-4 standard was originally published as IEC 801-4 in 1988 by IEC TC65 (Industrial Process Control). To apply this to all electronic equipment, IEC 1000-4-4 was published in January 1995, and from January 1997, 60000 was added so that ...
What is a fast transient/burst test?
The fast transient/burst test is a test that simulates noise caused by back electromotive force generated when interrupting inductive loads and noise caused by relay chattering, and evaluates the resistance of electronic equipment.