Bursts are applied at a repetition frequency of 5kHz or 100kHz.
Since burst pulses are not applied constantly, and depending on timing, they may not be applied at the point where a malfunction occurs, bursts are applied for 1 minute to apply them at various timings and increase the probability of detecting malfunctions.
If applied for a long time, bursts would be applied at various timings, so the probability of detecting malfunctions would increase, but it is thought that the standard set 1 minute or more as a round number that is neither too long nor too short.