[Test Equipment] Please tell me the capacitor capacity when outputting each waveform with the lightning surge tester LSS-15AX/F02/F03 series.
[Test Equipment] Please tell me the capacitor capacity when outputting each waveform with the lightning surge tester LSS-15AX/F02/F03 series.
・When outputting 1.2/50μF combination waveform from the surge out section: 10μF
・When outputting 1.2/50μF combination waveform from the power superimposition section: 10μF+2μF=12μF
・When outputting 10/700μF combination waveform: 10μF+2μF+8μF=20μF