Since the nut on the back of the panel must be secured before turning the pin, adjustment by the customer is not possible. (The wiring is secured with nuts, and there is a possibility of it breaking.) Please consult our Customer Service Center.
This function connects the waveform observation connector and coaxial cable attached to the CDN OUT or PULSE OUT of the tester, and checks whether pulses are being output normally from the tester. However, please note that this function is not ...
Simultaneous measurement is not possible. (When V and A are connected simultaneously, fluctuations occur in the ground potential, making it impossible to observe normal waveforms)
There are three possible causes listed below. If none of these apply, investigation by our Customer Service Center is required. ①The V monitor cable and A monitor cable are not connected to the oscilloscope simultaneously ②An isolation transformer is ...