Product Specifications (including optional items)
[Option] Up to what voltage can be applied to coupling clamp MODEL:15-00018A?
Application up to 5 kV is possible.
[Option] What cable diameter can be used with coupling clamp MODEL:15-00018A?
Approximately 37.5 mm.
[Option] What is the coupling capacitance of coupling clamp MODEL:15-00018A?
Approximately 50 pF to 100 pF.
[Test Equipment] What is the power supply voltage of the included power cable?
The standard included power cable is for AC100-240V. Please connect the power cable to a power outlet equipped with a protective ground terminal.
[Optional item] What is a radiation probe?
This is a probe used with an impulse noise simulator or fast transient/burst tester as the pulse generator, to apply a strong electric field to products, printed circuit boards, etc., for the purpose of identifying malfunction locations.
[Optional Accessory] What are the material, dimensions, weight, and load capacity of Test Table 03-00039A?
The material is wood, and the dimensions, weight, and load capacity are as follows. Dimensions: (W)1600×(D)800×(H)800mm Weight: Approx. 22kg Load capacity: Approx. 100kg
[Optional Accessory] What are the material, dimensions, and weight of Ground Plane 03-00007A?
The material is aluminum, and the dimensions and weight are as follows. Dimensions: (W)1800×(D)1000×(t)1.5mm Weight: Approx. 7.2kg (per plate) *03-00007A comes as a set of 3 plates.
[Optional Accessory] What is the screw size for connecting Ground Plane 03-00007A?
M3 size (with washer), 12 pieces are required for 3 plates. *12 pieces are included at the time of shipment.
[Optional item] What are the material, dimensions, mass, and load capacity of the insulating support stand 03-00024A?
The material is wood, and the dimensions, mass, and load capacity are as follows. Dimensions: (W) 1200 × (D) 1200 × (H) 1000 mm Mass: approx. 22 kg Load capacity: approx. 500 kg
[Optional item] What type of connector is used on the supplied coaxial cable?
Although there are coaxial connectors with a very similar appearance, this is the high-voltage coaxial connector “NMHV,” which we developed independently for use with test generators whose output pulse voltage exceeds 3 kV. Be sure to use the ...
[Test generator] Are there consumable parts for the fast transient/burst test generator?
The high-voltage relays and fuses inside the main unit are consumable parts, but their service life varies depending on the usage conditions and other factors.
[Test generator] How should the fuse of the fast transient/burst test generator FNS-AX4-A20/B63 be replaced?
There is a fuse holder in the AC inlet on the rear of the fast transient/burst test generator, and the customer can replace the fuse. If the recommended fuse is difficult to obtain, or if you have difficulty obtaining a fuse, please contact our ...
Does the remote software for FNS-AX4 support Windows 10 and 11 HOME/PRO?
It can be used with both Windows 10 HOME/PRO and Windows 11 HOME/PRO.
[Test generator] Why are PE and N rated at 10A for the EUT power capacity of FNS-AX4?
PE and N are the grounding point and neutral point, so ideally almost no current flows. If the load value of any phase changes, an unbalanced condition may occur, and a large current may also flow in the N phase. However, because such a condition is ...
[Test generator] The power capacity specification of FNS-AX4-B63 is three-phase AC600V 63A, with the N phase rated at 10A. For example, can a three-phase four-wire 32A EUT be tested?
If each phase is balanced (phase and voltage), no current flows in the N phase, so testing is possible. If the balance becomes extremely disrupted due to a failure or other cause, or if one phase is not output, more than 10A may flow in the N phase ...
[Test generator] What are the weight and dimensions of the fast transient/burst test generator FNS-AX4-A20/B63?
FNS-AX4-A20: External dimensions W430 × H222 × D437 (mm) (excluding protrusions); mass: approx. 14 kg FNA-AX4-B63: External dimensions W430 × H222 × D437 (mm) (excluding protrusions); mass: approx. 22 kg
[Test generator] Can the fast transient/burst test generator FNS-AX4-A20/B63 apply to any selected phase, rather than applying to all phases simultaneously?
In addition to simultaneous application to all phases, the FNS-AX4-A20/B63 can also apply to any selected phase. However, line-to-line testing requires the optional normal-mode coupling balun (15-00013A).
[Optional item] What is a coupling clamp?
It is a capacitive (electric-field) coupling clamp used when applying noise to signal lines. The standard requirements specify a length of 1000 mm and a height of 100 mm.
[Optional item] Up to how many kV can the coupling clamp (15-00012A) apply?
It can apply up to 5 kV.
[Optional item] Can the coupling clamp (15-00014A) be used for standard compliance testing?
15-00014A cannot be used for standard compliance testing. IEC 61000-4-4 specifies dimensional requirements for the coupling clamp.
[Tester] What is the power consumption of the fast transient/burst testers FNS-AX4-A20/B63?
Both FNS-AX4-A20/B63 consume approximately 120VA.
[Tester] Is there a warm-up time specification for the fast transient/burst testers FNS-AX4-A20/B63?
There is no warm-up time specification for the tester.
[Test Equipment] What happens when EUT power capacity exceeding the rating is input?
If voltage and current exceeding the rating are input, it may cause damage to the tester or fire. CDN components or wiring inside the tester may disconnect due to heat, so please use within the rated specifications.
[Tester] What is the difference between fast transient/burst testers FNS-AX4-A20 and FNS-AX4-B63?
FNS-AX4-A20 is a single-phase model and FNS-AX4-B63 is a three-phase model, so there is a difference in EUT power capacity. FNS-AX4-A20: Single-phase AC240V/20A DC125V/20A MAX FNS-AX4-B63: Three-phase AC600V/63A DC125V/63A MAX Please refer to our ...
[Tester] Please tell me about the outline specifications (features) of the fast transient/burst tester.
It complies with the IEC 61000-4-4 standard and can output burst noise to power lines and signal lines. The standard specifies common mode application (testing), but our company also enables normal mode application (testing) by using a coupling balun ...
How should the coupling clamp be used?
Place the coupling clamp on the ground plane and adjust it so that the coupling capacitance between the coupling clamp and the cable is maximized (so that the gap between the coupling clamp and the cable is minimized).
Is an isolation transformer required for fast transient/burst testing?
When connecting the AC power supply for the EUT to the tester, an isolation transformer is required for impulse noise testing, but it is not required for fast transient/burst testing.
Please tell me about the equipment arrangement (test environment).
The test environment for power lines involves placing an insulation support stand with a height of 0.1m on the ground plane, and placing the Equipment Under Test (EUT) on top of it. The length of the power line is 0.5m, or 1m for floor-standing ...
I would like to know the test overview of the fast transient/burst test.
Arrange the equipment according to the equipment arrangement example in the test room described in the standard, and conduct the test according to the test conditions (power lines and signal lines) and test levels. For signal line testing, conduct ...