ESS (Electrostatic Discharge Tester)
[Accredited Calibration] Is accredited calibration available for only the discharge gun + CR unit, or for the CR unit alone?
Accredited calibration of an electrostatic discharge simulator is performed only with the complete set (ESS main unit, discharge gun, applicable CR unit, and discharge cup). We cannot accept a unit if any component is missing. Calibration is ...
Is there a recommended calibration interval or calibration cycle for the electrostatic discharge simulators ESS-S3011A/B3011A?
The calibration interval or calibration cycle is to be determined by the customer, but we recommend calibration once a year.
The electrostatic discharge simulator ESS-S3011A has Normal mode and Extra mode. Does calibration of the tester need to be performed in both Normal mode and Extra mode? Or is calibration in only one of the modes sufficient?
Calibration in either Extra mode or Normal mode is sufficient. The reason is that the relay operation for generating the current waveform is the same in both modes, so the output current waveform is also the same.
The ISO 17025 calibration certificate for the electrostatic discharge simulator states: “The measured values of the current waveform are based on the average of 100 measurement results, and the current values are calculated by multiplying the voltage valu
For ISO 17025 accredited calibration, the work procedure is created by the applicant (our company) and carried out using a method approved by the accreditation body. Normally, the number of discharges is specified as 5 times in IEC61000-4-2 and 10 ...
Regarding ISO17025-accredited calibration for IEC61000-4-2, are there plans to discontinue support for Ed.2 after Ed.3 calibration begins?
There are currently no such plans. Ed.2 calibration will continue to be supported after Ed.3 calibration begins.
Is accredited calibration available for 18-00086B (voltage meter)?
It is not supported.