[Accredited Calibration] Is accredited calibration available for only the discharge gun + CR unit, or for the CR unit alone?

[Accredited Calibration] Is accredited calibration available for only the discharge gun + CR unit, or for the CR unit alone?

Accredited calibration of an electrostatic discharge simulator is performed only with the complete set (ESS main unit, discharge gun, applicable CR unit, and discharge cup). We cannot accept a unit if any component is missing.
Calibration is performed in accordance with IEC61000-4-2/ISO10605 Ed2/3. Combinations involving CR units not specified in the standards are not covered.
Measurement of a CR unit alone is not performed under accredited calibration.
(Under the manufacturer’s general calibration service, calibration of only the discharge gun, only the CR unit, and non-standard CR units is available.)