What is the pre-check function installed in the fast transient/burst tester FNS-AX4-A20/B63?
This is a function to check whether pulses are being output normally from the tester by connecting the waveform observation connector and coaxial cable attached to the CDN OUT or PULSE OUT of the tester. However, please note that this function is not a calibration of the tester.