Probe Stand MODEL: 03-00108A is marked as “not covered by IEC standards.” Does this mean that standards-compliant testing cannot be performed using this product?

Probe Stand MODEL: 03-00108A is marked as “not covered by IEC standards.” Does this mean that standards-compliant testing cannot be performed using this product?

Warning
In IEC 61000-4-2 testing, the general rule is that the gun is held by a person by hand during the test.
Therefore, testing with a fixed stand or robot, including ZAP (automatic electrostatic discharge application system), is outside the standard.
(In automotive ISO10605 Ed.3, there is an Annex stating that robots are permitted.)

Regarding whether testing using a fixture is NG, we believe there is no problem as long as the client and tester agree, such as by confirming the discharge current.
The test method must also be stated in the test report.