For VCP, peripheral equipment is installed to the left, right, front, and rear of the EUT, so application to at least four surfaces
is required.
For HCP, the electric field is considered to extend upward from the table surface and to be nearly uniform at every position,
allowing a uniform electric field to be applied to the bottom surface of the EUT.
Therefore, one application from the front of the HCP is the basic requirement.
If the EUT consists of multiple units, one application is made in front of each unit.
Application to the sides of the EUT is stated as “if possible,” so it is not mandatory (this follows the illustration in Figure 6).
In principle, one test point at the front is sufficient for the test to be effective.
Conversely, as shown in Figure 6 of Ed.3, one point on the side of a computer is also sufficient. (There is no requirement to apply the discharge to the long edge of the HCP.)