Standard
What is a lightning surge test under IEC standards?
It is a test that simulates surges generated by overvoltage from switching and lightning transients and evaluates the immunity (malfunction) of electronic equipment. Regarding lightning, it specifies induced lightning rather than direct lightning ...
What is the difference between lightning surge tests under IEC standards and JEC standards?
[IEC Standard] It is a test that simulates surges generated by overvoltage from switching and lightning transients and evaluates the immunity (malfunction) of electronic equipment. [JEC Standard] It is a test to verify the dielectric strength and ...
(Confidential) What is a Combination Waveform Generator circuit?
It is a circuit that produces both voltage and current surge outputs with a single circuit. When the output terminal is in an open circuit state, it outputs a specified voltage surge, and when the output terminal is in a short circuit state, it ...
(Confidential) The combination circuit of voltage surge 1.2/50μs and current surge 8/20μs is specified. Please explain the state during application (output).
IEC 61000-4-5 specifies two waveform verifications with a single generation circuit: voltage surge 1.2/50μs when open (open circuit) and current surge 8/20μs when short (short circuit). During testing, the waveform may deviate from the specified ...
In IEC61000-4-5 power line testing, the standard states "The non-linear current-voltage characteristics of the EUT must also be considered. Therefore, all lower levels including the selected test level must be tested." How should this be considered for an
Follow the reference standard, If not specified, conduct tests at all levels below and including the selected test level.
Is it possible to correlate tests between IEC and JEC lightning surge testers?
The equipment differences between IEC and JEC lightning surge equipment differ in generation method and effective output impedance, and the waveform behavior when impedance is connected to the output differs significantly, making correlation ...
Are there any grounding standard requirements for newly installing a lightning surge tester?
There is no specific grounding type requirement. However, please note that grounding is essential when conducting tests.
(Confidential) What is a combination waveform?
A combination waveform is a waveform output by a combination waveform generator circuit. There are mainly 1.2/50μ sec combination (1.2/50μ sec voltage surge, 8/20μ sec current surge) applied to power lines, and 10/700μ sec combination waveform ...
What is a combination waveform? What is a combination waveform generator circuit?
A combination waveform is a waveform output by a combination waveform generator circuit. There are mainly 1.2/50μs combination (1.2/50μs voltage surge, 8/20μs current surge) applied to power lines, and 10/700μs combination waveform (10/700μs voltage ...
What are the main JEC standard numbers that specify the lightning surge test?
For impulse waveforms, there is mainly JEC-0202 (1994), which was revised from JEC-212 (1981). For impulse withstand voltage test methods, there is mainly JEC-0103 (2005), which was revised from JEC-210 (1981). JEC-0202 (1994) has been consolidated ...
What is the IEC 61000-4 series standard number that specifies the lightning surge test?
The lightning surge test is specified in IEC 61000-4-5.
(Confidential) What is the lightning surge test?
This is a test that simulates surges generated by overvoltage from switching and lightning transients to evaluate the immunity (malfunction) of electronic equipment. For lightning, it specifies induced lightning rather than direct lightning strikes, ...