Differences in test results may occur when the discharge gun is different. The waveforms of TC and GT are different, so they are not completely equivalent. Other factors include the angle of the discharge gun when discharging to the EUT, the routing ...
The first peak specified for the current waveform is formed by the parasitic capacitance between the discharge gun and the mounting plate. Because the simplified version has a smaller area than the IEC standard product, the first peak will be ...
Let X be the value observed on the oscilloscope (unit: V). X × 10 (20dB) × 10 (20dB) ÷ 2 (2Ω resistance of the target) = peak current value (unit: A). Example: If the oscilloscope displays 0.15V when 2kV is applied (The standard specifies 7.5A ±15% ...
It is unlikely that discharge would be restored by wiping and cleaning the gun tip (discharge tip). The purpose of cleaning is to maintain the insulation performance of the resin housing. If contamination or other factors reduce insulation ...
To perform more reliable testing, each tester is equipped with the following pre-check functions: ESS-S3011A: ①High-voltage power supply output check, ②Insulation failure check, ③Discharge gun discharge relay check ESS-B3011A: ①High-voltage power ...