Product Specifications (including optional items)
The LANGER near-field probes RF and LF1 use SMA-BNC, whereas XF1 uses SMA-SMA. Is there a reason for this?
XF1 (30MHz to 6GHz) uses SMA-SMA because of its high-frequency characteristics. BNC connectors support frequencies up to approximately 3GHz. *Reference: The conversion connector for connecting 00-00014A to XF1 is 00-00137A (N(P)-SMA(J)). Reference ...
What equipment is required for the EPS-02Ev3 spatial electromagnetic field visualization system?
The required equipment (system configuration) consists of a PC that controls the entire system (including visualization software), a spectrum analyzer, an electromagnetic field probe (including a probe cover head), a camera, and a preamplifier.
How does the camera detect position?
A color-recognition probe cover head is fitted to the tip of the electromagnetic field probe, and the camera detects its position based on the color of the probe cover head. Probe cover heads are supplied in three colors (yellow, red, and blue).
There are spatial electromagnetic field visualization systems other than EPS-02Ev3. What are the differences?
EPS-02Ev3 has a measurement frequency range of 100kHz to 3GHz (depending on the electromagnetic field probe specifications), while EPS-02EMFv2 and EPS-02Hv2 are designed for low-frequency magnetic field measurement (10Hz to 400kHz). EPS-02S is also ...
Which probes can be used with the EPS-02Ev3 spatial electromagnetic field visualization system?
Electro-Metrics EM-6992 electromagnetic field probe set is offered as standard, but LANGER near-field probes and other probes can also be used. Please feel free to contact us if you would like a demonstration using your own probe.
I own an EPS-02Ev2 spatial electromagnetic field visualization system. What has been changed (added) in EPS-02Ev3?
The major added features are density display, three-dimensional display (time, frequency, and intensity), and difference display.
Can the EPS-02Ev2 spatial electromagnetic field visualization system be upgraded to EPS-02Ev3?
Yes, it can be upgraded. Please contact us separately for detailed specifications and a quotation.
Please tell me the resolution of the probe image recognition on the EPS-02Ev3.
The minimum resolution is approximately 3 mm.
Please explain how to capture (record) data at measurement points with the EPS-02Ev3 spatial electromagnetic field visualization system.
SINGLE / FREE RUN/ MAX PEAK DATA /MAX HOLD to capture (record) data. SINGLE:Measures only once. FREE RUN:Continuously overwrites the data. MAX PEAK DATA:Compares the maximum peak value of the spectrum waveform and overwrites the data when the stored ...
Please tell me the PC specifications required for the EPS-02Ev3 spatial electromagnetic field visualization system.
OS: Microsoft Windows 10(Japanese / English versions)Microsoft Windows 11(Japanese / English versions) CPU: Intel Core i5 or higher (i7 or higher recommended) Memory: 8 GB or more recommended Storage: At least 10 GB of free space Display: WXGA ...
The probe head cover has torn. Can I purchase a replacement?
Yes, it is available for purchase. Please purchase the replacement three-color probe cover head (03-00123A).
For EPS-02Ev2 setup, I installed NI-VISA, but during installation the error message “A third-party VISA library was detected” appeared. Is it safe to continue installing NI-VISA?
A VISA library that conflicts with NI-VISA may already be installed. If this will not cause issues with other applications, uninstall the third-party VISA library before installing NI-VISA.
When using PN7405 series probes (manufactured by ETS-Lindgren), are there any precautions regarding the PN7405 antenna factor included as standard with EPS?
The antenna factor created by our company is a corrected value that subtracts 41dB for the preamplifier (00-00019A) in addition to the probe. Therefore, if another preamplifier is used or no preamplifier is used, select through or edit and create a ...
During measurement, the spectrum value on the EPS software screen differs from the value displayed on the spectrum analyzer by as much as approximately 40dBμV at some frequencies. Why?
Is a MODEL for one of the antennas selected under Antenna Factor in the EPS software? If so, the value displayed on the software screen is corrected using the antenna factor. Clear the “Enabled” checkbox for Antenna Factor or select “THROUGH”, and ...
For the frequency response data of the ETS-PN7405 (manufactured by ETS-Lindgren), should the ETS-Lindgren data or the Noise Laboratory data be used with EPS?
ETS-Lindgren performs calibration using the TEM-CELL method (entire space), while Noise Laboratory performs calibration using a stripline (close proximity). Considering the intended use with EPS, stripline calibration is closer to actual measurement ...
Is measurement in QP mode possible with EPS-02Ev3?
EPS-02Ev3 does not have a QP mode measurement function, and there are no plans to add one. This is because it would increase the measurement time.
After changing the PC, the spectrum analyzer (FSV) can no longer be used. How should this be handled?
Check that the software IP address matches the PC address, and also check the connection.
The antenna factor supplied as standard with EPS-02Ev3 is data acquired as a set comprising the probe, coaxial cable, and preamplifier. When recreating a factor for the probe alone, should the gain of the coaxial cable and preamplifier be subtracted or ad
Subtract the gain attributable to the coaxial cable and preamplifier.
Can the display unit for measurement data in the EPS-02Ev3 software be set to [A/m]?
The EPS-02Ev3 software does not have a setting to switch the display unit for measurement data to [A/m]. EPS-02Ev3 performs measurements near the EUT. In the near-field region, the ratio between the electric and magnetic fields varies significantly ...
When installing EPS-02Ev3, National Instruments software must also be installed, but it appears that unnecessary software is installed as well. Please tell me which software is required to operate EPS-02Ev3.
The following three NI products are the minimum required to operate EPS-02Ev3. ? 53 NI Update Service ? 82 NI-VISA ? 92 NI-VISA Runtime If the measuring instrument is connected using a GPIB-USB cable, the following three products are also required in ...